You are viewing a free preview of this lesson.
Subscribe to unlock all 14 lessons in this course and every other course on LearningBro.
Spec mapping: OCR H556 Module 6.5 — Medical Imaging (the Doppler effect applied to reflected ultrasound; the two-way reflection geometry doubles the single-shift formula to Δf/f0≈2vcosθ/c for a reflector moving at velocity v at angle θ to the beam; measurement of blood-flow velocity in arteries and veins; colour-Doppler imaging; the angle-dependence trade-off in clinical practice). Refer to the official OCR H556 specification document for exact wording.
Pulse-echo ultrasound, as described in the previous lesson, tells you where a reflecting boundary is — it reads off the depth from the round-trip pulse time. Doppler ultrasound goes one step further: it measures how fast that boundary is moving towards or away from the transducer. By applying the technique to the red blood cells flowing in an artery or vein, clinicians can measure blood-flow velocity non-invasively and in real time. The diagnostic implications are enormous: narrowed arteries, leaky or stenotic heart valves, foetal heart activity, deep-vein thrombosis and the perfusion of transplanted organs are all routinely assessed by Doppler ultrasound.
This lesson covers the Doppler-ultrasound section of Module 6.5 — Medical Imaging of the OCR A-Level Physics A specification (H556) — the physics of the frequency shift, the central formula Δf/f0≈2vcosθ/c, the geometry, clinical applications and the colour-Doppler imaging mode.
You met the Doppler effect earlier in the course, in the context of audible-frequency sound: when a source or observer moves relative to a medium, the observed frequency differs from the source frequency. The same physics applies to ultrasound — and with one important extra twist when the wave is reflected by a moving target rather than directly emitted by a moving source.
Consider a stationary transducer emitting ultrasound at frequency f0. A reflector (a red blood cell, say) moves at velocity v along the beam direction. There are two Doppler shifts in series:
The net shift detected at the transducer is therefore approximately twice the single-stage shift:
f0Δf≈c2vcosθ
where c is the speed of sound in the medium (≈1540 m s−1 in soft tissue, ≈1560 m s−1 in blood) and θ is the angle between the ultrasound beam and the direction of motion of the reflector. The factor of 2 is the signature of a reflected-wave Doppler measurement — it is not present in a one-way Doppler shift (e.g. the redshift of a distant galaxy).
Rearranging for the velocity:
v=2f0cosθcΔf.
This is the central A-Level Doppler-ultrasound equation. Measuring Δf (the difference between emitted and received frequencies) and knowing f0, c and θ, the scanner computes v.
The factor of cosθ is essential. The reflector's velocity v has a component vcosθ along the beam direction, and only this component produces a Doppler shift; the perpendicular component does not change the round-trip path length and so does not shift the frequency.
Because vcosθ is what is actually measured, errors in estimating the angle translate directly into errors in the reported velocity. Ultrasound scanners display a steering-line overlay on the B-scan image that the operator aligns with the vessel axis; the software then knows θ and computes v from Δf.
The diagram shows the central idea: the beam is incident at angle θ to the flow direction; only the component vcosθ along the beam contributes to the Doppler shift; the reflected wave returns to the transducer with a wavelength shorter (and frequency higher) than the emitted wave when the reflector is moving towards the transducer.
An ultrasound beam of frequency 5.0 MHz is aimed at a blood vessel at an angle of 60° to the direction of flow. The observed Doppler shift is 300 Hz. Take the speed of sound in blood as c=1560 m s−1. Calculate the blood-flow velocity.
Solution.
v=2f0cosθcΔf=2(5.0×106)(cos60°)(1560)(300)=2(5.0×106)(0.5)4.68×105=5.0×1064.68×105≈0.094m s−1.About 9.4 cm s−1 — well within the normal range for venous blood flow.
The blood in a patient's carotid artery is flowing at v=0.60 m s−1. A 4.0 MHz ultrasound beam is incident at 45° to the direction of flow. Take c=1560 m s−1 in blood. What Doppler shift is observed?
Solution.
Δf=c2f0vcosθ=15602(4.0×106)(0.60)(cos45°)=15602(4.0×106)(0.60)(0.707)≈2175Hz≈2.2kHz.A shift of about 2 kHz. Doppler shifts in clinical ultrasound typically range from ∼100 Hz (slow venous flow) to several kHz (fast arterial flow, especially near a narrowing). The shift falls within the audible range — and most Doppler scanners play it through a loudspeaker so the operator can hear the flow.
In Worked Example 1, an operator misreports the beam–flow angle as 50° when the true angle is 60°. What is the percentage error in the reported velocity?
Solution. The reported velocity is vreported=cΔf/(2f0cos50°), while the true velocity is vtrue=cΔf/(2f0cos60°). The ratio is
vtruevreported=cos50°cos60°=0.6430.500≈0.778.
The reported value is about 22% low. Conversely, an over-estimate of θ produces an over-estimate of v. Doppler-ultrasound velocity measurements are inherently sensitive to angle, especially near θ→90° where cosθ→0 and the formula diverges. This is why scanners refuse to compute v at θ>70° and why clinicians keep the beam at 45°–60° wherever possible.
Doppler ultrasound is routinely used to:
Modern scanners combine Doppler with B-scan imaging to produce colour-Doppler displays, in which the velocity is shown as a colour overlay on the grey-scale anatomical image: red for flow towards the transducer, blue for flow away. This makes vascular anatomy vivid and intuitive — and shows turbulent or reversed flow as a multi-colour mosaic, instantly highlighting clinically interesting regions.
flowchart TB
Q["Ultrasound question:<br/>position or velocity?"]
Pos["Position<br/>(structure)"]
Vel["Velocity<br/>(motion)"]
PE["Pulse-echo<br/>A-scan, B-scan<br/>d = ct/2"]
Dop["Doppler<br/>Δf/f₀ = 2v cosθ/c"]
Col["Colour Doppler<br/>(B-scan + Doppler overlay)"]
Q --> Pos
Q --> Vel
Pos --> PE
Vel --> Dop
Dop --> Col
The choice is clean: pulse-echo for static structure, Doppler for moving reflectors, and colour Doppler when you want both at once.
Question (8 marks): A vascular sonographer uses Doppler ultrasound at f0=4.0 MHz to measure the blood-flow velocity in a patient's carotid artery. The beam is incident at an angle of θ=60° to the vessel axis. The speed of sound in blood is c=1560 m s−1. The observed Doppler shift is Δf=1500 Hz.
(a) Derive (or state, with reason) the Doppler-shift formula Δf/f0=2vcosθ/c for ultrasound reflected from a moving target. Explain why the formula contains a factor of 2 that is absent from a one-way Doppler shift. [3]
(b) Use the data to calculate the blood-flow velocity. [2]
(c) The sonographer realises she has measured the angle wrong; the true angle is 50°, not 60°. Calculate the corrected velocity, and comment on the magnitude of the error. [3]
Subscribe to continue reading
Get full access to this lesson and all 14 lessons in this course.